End thermal resistance
- Tel:0550-7591188
- Fax:0550-7591188
- Email:2136705934@qq.com
Description
The end thermal resistance element is made of specially processed wire and is attached to the front end of the thermometer. Compared with the general axial thermal resistance, the end thermal resistance can reflect the actual temperature of the measured end more accurately and quickly, and is suitable for measuring the end temperature of bearing bush or other parts.
● Main technical parameters
Temperature sensing element 0℃ resistance value (R100) and its resistance at 0℃ R0 classification number Cu50 :R0= 50±0.050 ω Cu100 :R0=100± 0.10 ω classification number Pt100:R0=100±0.12 ω (grade B) where :R0 is the component at 0℃ resistance value.
● Measuring range and tolerance (unit: mm)
The name of the | Temperature range: ℃ | Allowable deviation △t℃ | Degree, |
End copper resistor | - 50-100 | Plus or minus (0.3 + 6.0 x 10-3 | t |) | Or Cu100 Cu50 (G) |
End platinum resistor | - 150-200 | B: - (0.30 + 0.005 | t |) | Pt100 |
specifications
Type no. | Degree, | Structural features | Temperature range: ℃ | Protective tube material | Thermal response time | specifications | Nominal pressure | External wiring lead length L (mm) | |
Diameter D (mm) | M (mm) | ||||||||
WZCM-201 | Cu50 Cu100 | Fixed embedment or threaded connection | - 50-100 | Copper T2 - Y | The < 15 | Φ 6 | M8x0 .75 | Atmospheric pressure | 500 |
WZPM-201 | Pt100 | - 150-200 | The < 15 | Φ 8.7 | M10x1 | ||||
WZPM-201B |
Note: WZPM-201B adopts the imported component WZPM-018, the accuracy grade is A.